Sirona Dental A/S 4Installation
Operating Instructions DAC UNIVERSAL 4.2Installation of DAC UNIVERSAL connections
båÖäáëÜ
4.2.3.1
Permanent water connection
The following must be observed when connecting the DAC UNIVERSAL
to a permanent water supply:
Automatic al water connection _Sirona without Ja pan_Nitram
To operate the DAC UNIVERSAL with a water conditioning unit, it must
be switched to "Automatic water supply". This is done via the menu
system, see "Menu system DAC UNIVERSAL ADVANCED [ 77]" or
"Menu system DAC UNIVERSAL STANDARD [ 165]“.
Empty was te water tank_Sirona without Japan
If you use a permanent water supply and have not installed a siphon,
please empty the waste water tank once a day or whenever it is full, see
"Remove and clean the drain tank [ 131]".
4.2.3.2
Manual filling of the water tank
If no permanent water connection is selected, the integrated water tank
must be filled with water manually.
See "Filling the water tank [ 69]" for the procedure.
4.2.4 Waste water connection
The DAC UNIVERSAL works only in a one-way water system, i.e. used
water (waste water) is not reused.
Therefore, the "Drain" connection on the back of the DAC UNIVERSAL
must be connected either to:
A waste water tank, see "Connecting the waste water tank [ 52]" or
A siphon, see "Connecting the siphon [ 52]"
IMPORTANT
Observe local requlations for backflow protection when connecting the
DAC UNIVERSAL to permanent water supply. According to EN1717 an
air gap is required as backflow protection of Category 5 water."
NOTICE
A water filter (20 µm) must be installed upstream of the DAC
UNIVERSAL to ensure that the water is clean.
NOTICE
Do not start a cycle if the DAC UNIVERSAL is not connected to the drain
tank or drainage system.
IMPORTANT
Dispose of the waste water according to the national guidelines and the
applicable regulations for waste disposal.