60 53 362 D3472
24 D3472.201.10.20.02 07.2016
3Product description Sirona Dental A/S
3.1Functions Operating Instructions DAC UNIVERSAL
Sterilization
The wrapped instruments are sterilized for 3 minutes at 134°C (273°F).
Functions WRAPPED lid with ADVANCED
For instructions for the sterilization cycle for wrapped instruments see
"Reprocessing with the WRAPPED lid DAC UNIVERSAL ADVANCED
[ 110]" for the DAC UNIVERSAL ADVANCED and "Reprocessing with
the WRAPPED lid DAC UNIVERSAL STANDARD [ 182]" for the DAC
UNIVERSAL STANDARD.
3.1.4 Functions with the FLEX lid
The unwrapped FLEX lid instruments are reprocessed as follows using
the appropriate adapter on the FLEX lid.
Internal cleaning
Internal cleaning ensures that the water and air channels of the
instruments are rinsed thoroughly. For the internal cleaning of the
instruments, water is forced through the channels using compressed air.
The water and air channels are rinsed with water 12 times during a
cleaning cycle.