Sirona Dental A/S 5Operation DAC UNIVERSAL ADVANCED
Operating Instructions DAC UNIVERSAL 5.3Reprocessing with the DAC UNIVERSAL ADVANCED
båÖäáëÜ
5.3.5 Reprocessing with the FLEX lid DAC UNIVERSAL
ADVANCED
The FLEX lid is used to clean and thermally disinfect ultrasonic scaler
handpieces, scaler tips, airflow nozzles and sprayvit nozzles.
Tip: If there are many instruments to be decontaminated or to ensure the
instruments will be available again quickly, we recommend using a
second FLEX lid.
5.3.5.1
General placement of instruments DAC UNIVERSAL ADVANCED
General attachment
Place the instruments on their respective adapters of the FLEX lid.
Depending on the instrument and the kind of adapter, you must press or
click the instrument onto the adapter, activate a snap lock on the adapter,
or tighten a threaded ring on the instrument.
Please make sure that you use the lever on the side of the adapter (if
applicable) and hear a slight "click" when you place the instrument on the
adapter.
Pull lightly on the instrument to ensure that it is firmly attached to the
adapter.
IMPORTANT
The FLEX lid is only for cleaning and high-level disinfection.
NOTICE
All instruments must be placed straight onto the respective adapter to
prevent damage to the O-rings.
NOTICE
The instruments must be easy to attach to the adapters. If the instrument
does not attach to the adapter easily, it can be damaged.
The manufacturer does not accept responsibility for instruments that are
damaged due to being subjected to force.
WARNING
If instruments are attached to an unsuitable adapter or if instruments are
damaged, disinfection and cleaning may fail. In this case, reprocessing
of the instruments must be repeated.