60 53 362 D3472
116 D3472.201.10.20.02 07.2016
5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
General attachment of scaling tips
1. The scaler tip was unscrewed from the ultrasonic handpiece with the
torque wrench and the tip remains in the torque wrench.
2. Ensure that the torque wrench is correctly attached to the tip
General placement
The instruments should not be leaning on each other; otherwise they can
be attached as preferred by the user.
Placement is described in the separate sections if it deviates from the
general placement described here.
See "Adapters for the FLEX lid [ → 238]" for detailed instructions on the
individual adapters.
WARNING
Scaler tips must be inserted with the associated torque wrench. The
torque wrench should remain on the tip during the whole process.
NOTICE
Tips have to be screwed straight onto the adapter so they can’t jam up.