Sirona Dental A/S 4Installation
Operating Instructions DAC UNIVERSAL 4.1Requirements of the location / Ambient conditions
båÖäáëÜ
4
Installation
4.1
Requirements of the location / Ambient conditions
Requirements To find a suitable location for the DAC UNIVERSAL, observe the
following requirements:
● ● Open, well-ventilated area
● Position in the unclean area directly adjacent to the clean area in the
hygiene room
● Level, heat-resistant countertop
● Close to a compressed air and power source
● Recommended minimum distance from the wall: 10 cm
● Recommended minimum height: 70 cm
If the DAC UNIVERSAL needs to be switched off completely, the power
plug must be accessible.
Ambient conditions The DAC UNIVERSAL may be used only indoors under the following
ambient conditions:
● Relative humidity from 60% to max. 80%
● Up to an altitude of 2,000 m above sea level
● Room temperature between 16°C and 26°C (61°F and 79°F)
● Fluctuations in electrical supply should remain within ± 10% of the
nominal voltage
● Transient overvoltage according to installation category II
● Pollution degree 2
4.2
Installation of DAC UNIVERSAL connections
CAUTION
Risk of crushing
The recommended minimum height for installing the DAC UNIVERSAL
is 70 cm, in order to prevent injuries when removing the lid.
IMPORTANT
Only technicians trained by the manufacturer with a training certificate
issued by the manufacturer are qualified to install, service, repair, and
maintain the DAC UNIVERSAL.
IMPORTANT
Please note that the DAC UNIVERSAL must be warmed up to room
temperature before being connected.