EasyManua.ls Logo

Sirona DAC UNIVERSAL - Reprocessing with the WRAPPED lid DAC UNIVERSAL ADVANCED

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
110 D3472.201.10.20.02 07.2016
5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
5.3.4 Reprocessing with the WRAPPED lid DAC UNIVERSAL
ADVANCED
The WRAPPED lid is used for reprocessing wrapped rotating handpieces
and wrapped solid instruments.
WRAPPED lid not for F LEX lid instruments
IMPORTANT
The WRAPPED lid can not be used for the FLEX lid instruments. To get
FLEX lid instruments sterilized, a class B sterilzer (e.g. DAC
PROFESSIONAL) is necessary.
IMPORTANT
The WRAPPED lid is intended only for sterilizing wrapped instruments.
Do not reprocess any unwrapped instruments in the WRAPPED lid.
WARNING
Before using the WRAPPED lid, the sterile filter included with the unit
must be installed by a qualified service technician.
IMPORTANT
See "Functions with the WRAPPED lid [ 23]" for compatibility and the
exact requirements of the WRAPPED lid.
IMPORTANT
Packing-/wrapping material used with the WRAPPED lid in the DAC
UNIVERSAL must be in accordance with EN868. Sealable paper bags
and sealable transparent bags or tubes made of porous material and
plastic composite film must be filled to only 3/4 of their total volume.
IMPORTANT
Overfilling may strain the seams.
There must be at least 30 mm space underneath the sealing rim
between the sterilized product and the seam to be sealed
WARNING
If the wrapping of the wrapped instruments has been damaged, the
instruments needs to be rewrapped and resterilized.

Table of Contents

Other manuals for Sirona DAC UNIVERSAL

Related product manuals