9Cleaning and maintenance DAC UNIVERSAL STANDARD Sirona Dental A/S
9.10Cleaning the empty chamber of the DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
6. Press the "Start"-Key.
A prompt to press down on the safety ring appears on the display
(
"Press safety ring!"
).
7. Press down on the safety ring to start the cycle. If the safety ring is
not pressed within 20 seconds, you must choose the lid and press the
"Start"-Key again.
The STANDARD lid of the DAC UNIVERSAL closes as soon as
the safety ring is pressed.
The DAC UNIVERSAL automatically runs the cleaning and
sterilization cycle.
You can see on the display at any time which phase the DAC
UNIVERSAL is in.
The following message appears on the display when the selected
cycle is finished:
"Press C to unload"
. The lid moves about
halfway out of the chamber. This indicates that a cycle has been
completed.
8. Press the "Clear"-Key.
The STANDARD lid will open completely.
9. Remove the STANDARD lid from the holder on the DAC UNIVERSAL
at the end of the cycle.
10. Place the STANDARD lid on the counter or put it in the optional lid
holder.
The chamber and hoses are clean.
IMPORTANT
If no program is selected, the standard program is used.
CAUTION
Danger of burning!
Never hold your head directly over the DAC UNIVERSAL when it opens
and do not hold the lid firmly, as steam comes out of the unit when it
opens.
CAUTION
Danger of burning!
The metal part of the lid is very hot. Please touch only the plastic part of
the STANDARD lid at the end of the cycle.