Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD
båÖäáëÜ
5. Insert the FLEX lid in the cover holder of the DAC UNIVERSAL in
such a way that the instruments are pointing down.
There are instruments on the FLEX lid; the DAC UNIVERSAL is
still open and ready for a cycle.
6. Press the "Menu/Enter"-Key twice and select the desired program
using the arrow keys.
7. Confirm your input by pressing the "Menu/Enter"-Key again.
8. Press the "Clear"-Key.
Depending on the program selected, the following message may
appear on the display:
"Ready 134 °C"
,
"Ready 121 °C"
or
"Ready 134 °C 18min"
.
9. Press the "Start"-Key.
A prompt to select the lid will appear on the display.
10. Select the FLEX lid using the arrow keys.
11. Press the „Start“-Key twice.
On the DAC UNIVERSAL, only the LED at the left of the "Rinse/
Flush"-Key is on.
A prompt to press down on the safety ring appears on the display.
12. Press down on the safety ring to start the cycle. If the safety ring is
not pressed within 20 seconds, you must choose the lid and press the
"Start"-Key again.
The FLEX lid of the DAC UNIVERSAL closes as soon as the
safety ring is pressed.
The DAC UNIVERSAL automatically runs the cleaning and high-
level disinfection cycle.
You can see on the display at any time which phase the DAC
UNIVERSAL is in.
The following message appears on the display when the selected
cycle is finished:
"Press C to unload"
. The lid moves about
halfway out of the chamber. This indicates that a cycle has been
completed and the instruments have been reprocessed.
NOTICE
Make sure that the lid is inserted all the way into the holder so that the
instrument heads do not touch the wall of the chamber. Make sure there
are no objects in the chamber.
NOTICE
The DAC UNIVERSAL and/or the instruments can be damaged if a lid
is not inserted correctly.
WARNING
If no program is selected, the standard program for reprocessing is
used.