8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.3Reprocessing with DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
13. Press the "Clear"-Key.
The FLEX lid will open completely.
14. Remove the FLEX lid from the holder on the DAC UNIVERSAL at the
end of the cycle to allow the instruments to cool down faster.
15. Place the FLEX lid on the counter or put it in the optional lid holder.
16. Check the instruments visually to make sure they are dry.
17. Check the instruments visually to make sure they have been cleaned.
18. Check the cycle number, properly conducted cycle on your
documentation (printer, software, etc.): For for high-level disinfection,
"High-level disinfection, time, ok" must be indicated.
19. Check whether the dark status bar of the chemical indicator is in the
"ACCEPT" range.
If everything is correct, reprocessing was successful.
20. Document the cycle according to the work instructions applicable in
your practice.
CAUTION
Danger of burning!
Never hold your head directly over the DAC UNIVERSAL or touch the
inserted lid, while the DAC UNIVERSAL opens, as steam comes out of
the unit upon completion of a cycle.
WARNING
After completion of the cycle, make sure that all instruments are still
properly attached to the FLEX lid or the adapters. If instruments have
fallen into the chamber during the cycle or if an instrument is no longer
properly attached after the end of a cycle, cleaning and high-level
disinfection have failed. In this case, reprocessing of the instruments
must be repeated.
CAUTION
Danger of burning!
Even when the FLEX lid is half or completely open, the instruments are
still hot. Allow the extremely hot instruments to cool down after the
process.
CAUTION
Danger of burning!
The metal part of the lid is very hot. Please touch only the plastic part of
the FLEX lid at the end of the cycle.
IMPORTANT
When they are cooling, the heads of the instruments should point
upwards to prevent corrosion.