Operating Instructions DAC UNIVERSAL 8.3Reprocessing with DAC UNIVERSAL STANDARD
båÖäáëÜ
21. Remove every instrument from the FLEX lid. See "Attaching
instruments to the respective FLEX lid adapters [ → 238]" for detailed
instructions on removing the individual instruments.
The instruments can now be used again.
High-level disinfection of unwrapped instruments is not sufficient for
invasive procedures. Instruments must be wrapped and a sterilization
must be conducted in a class B sterilizer (e.g. DAC
PROFESSIONAL).
After the end of the cycle, the DAC UNIVERSAL automatically
reactivates lubrication for the next cycle.