Sirona Dental A/S 8Operation DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 8.2Using the menu and display information DAC UNIVERSAL STANDARD
båÖäáëÜ
"Settings" "Expel"
0 - 3 Purging Purge instruments with air after sterilization.
Standard setting: 0
Can be incresed to level 3.
Level 0: no blowing
Level 1: 50 sec. effective blowing time
Level 2: 100 sec. effective blowing time
Level 3: 150 sec. effective blowing time
Please note that increasing the blowing time prolongs the
total cycle time.
Level 1 is recommended for instruments of the manufacturers
W+H, KaVo, NSK, and most other instruments. Level 3 is
recommended for Sirona T1 CLASSIC/TE. However, please
observe the manufacturer's instructions for use for each
instrument.
If processing instruments for invasive surgical treatments,
Expel function cannot be used without prior installation of a
sterile-filter.
"Set time/date" "Set year"
(00-
99)
Setting the year.
"Set month"
(01-12)
Setting the month.
"Set date"
(01-
31)
Setting the day.
"Time display
unit"
(12h / 24 h)
Setting the time display
"Set hour"
(01-
24)
Setting the hours.
"Set min"
(00-59 Setting the minutes.
Menu level 1 Menu level 2 Menu level 3 Result / Activity / Note