8Operation DAC UNIVERSAL STANDARD Sirona Dental A/S
8.2Using the menu and display information DAC UNIVERSAL STANDARD Operating Instructions DAC UNIVERSAL
"Settings" "Automatic
water"
"ON "
Automatic filling of the water tank by a water conditioning
system connected to the DAC UNIVERSAL.
"OFF"
Manual filling of the water tank.
"Oil Settings"
1-3 oil settings Oil quantity for lubrication of instruments.
Standard setting: Level 1.
Can be increased up to level 3.
Level 1: standard.
Level 2: 25% more than level 1.
Level 3: 50% more than level 1.
"Ambient
Sensitivity"
0-4 ambient
temperature
sensitivity
Standard setting: Level 1 (low ambient temperature
sensitivity).
Level 0: ambient temperature sensitivity deactivated.
Level 1: Low ambient temperature sensitivity.
Level 2: Medium ambient temperature sensitivity.
Level 3: High ambient temperature sensitivity.
Level 4: Maximum ambient temperature sensitivity.
When there is a low temperature in the practice/clinic, we
recommend increasing the ambient temperature sensitivity
somewhat – see "Requirements of the location / Ambient
conditions [ → 49]".
Please note that activating or increasing the level of sensitivity
can potentially prolong the total cycle time and increase
power consumption.
"History" "Cycles"
Shows the number of cycles completed.
"Errorbuffer"
Shows the latest messages. They can be selected using the
"Up" or "Down"-Key.
"Service" Is displayed when the next service of the unit is due. This is
indicated by the number of remaining cycles until the next
service and as a date (mm/dd/yyyy) of the next service.
"About"
Displays the serial number of the unit and the software
version.
Menu level 1 Menu level 2 Menu level 3 Result / Activity / Note