60 53 362 D3472
120 D3472.201.10.20.02 07.2016
5Operation DAC UNIVERSAL ADVANCED Sirona Dental A/S
5.3Reprocessing with the DAC UNIVERSAL ADVANCED Operating Instructions DAC UNIVERSAL
22. Remove every instrument from the FLEX lid. See "Attaching
instruments to the respective STANDARD lid adapters [ 220]" for
detailed instructions on removing the individual instruments.
The instruments can now be used again.
High-level disinfection of unwrapped instruments is not sufficient for
invasive procedures. Instruments must be wrapped and a sterilization
must be conducted in a class B sterilizer (e.g. DAC
PROFESSIONAL).