EasyManua.ls Logo

Sirona DAC UNIVERSAL - Page 215

Sirona DAC UNIVERSAL
248 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 53 362 D3472
D3472.201.10.20.02 07.2016
215
Sirona Dental A/S 10Display messages DAC UNIVERSAL STANDARD
Operating Instructions DAC UNIVERSAL 10.1Display messages and what they mean DAC UNIVERSAL STANDARD
båÖäáëÜ
96 Too little pressure during
purging of instruments.
Please contact your authorized dealer.
97 No water during external
cleaning.
Press the "Clear"-Key to reset and then restart the DAC UNIVERSAL.
If the water level is too low, fill demineralized or distilled water (water
conductivity < 3.0 µS/cm) into the water tank.
Press the "Rinse/Flush"-Key until water appears at the bottom of the
chamber. If this is the case, see "Drain the water tank of the DAC
UNIVERSAL STANDARD [ 196]".
If no water appears in the chamber, please contact your authorized dealer.
Empty the drain tank if it is full.
Clean or change the drain filter.
Make sure the drain hose is not blocked, defective, or kinked.
IMPORTANT: The drain hose must be made of a heat-resistant material
such as PTFE.
If the error occurs repeatedly, please contact your authorized dealer.
98 The lid will not close. Press the "Clear"-Key to reset and wait until the lid is in its highest position.
Make sure that the lid was placed properly in the lid holder.
Look for objects that prevent the lid from closing completely and remove
them.
Start a new cycle.
If the error occurs repeatedly, please contact your authorized dealer.
99 The program settings must be
checked.
Please contact your authorized dealer.
Code
no.
Description Corrective action

Table of Contents

Other manuals for Sirona DAC UNIVERSAL

Related product manuals