60 53 362 D3472
42 D3472.201.10.20.02 07.2016
3Product description Sirona Dental A/S
3.5Presentation of the sterilization process Operating Instructions DAC UNIVERSAL
Sterilization process FLEX lid
FLEX lid
The following graph shows temperature (°C) and pressure (bar) during
the standard program with the FLEX lid, with standard settings, holding
time 3 min and a total cycle time of 22 min:
FLEX lid
1.A Pressure test with air
2.B Internal flush of the channels with water
3.D 5x External cold rinse
5.E 1x External warm rinse
4.F Drain chamber
6.G Back-Flush of the channels with steam
7.H Disinfection
8.I Ventilation of the chamber, open lid

Table of Contents