Step 54 Set a Tidal volume of 8ml and a Max PIP of 30mbar
Verify that:
The displayed waveform changes
showing a rise in flow with a sharp
cut off.
Step 55 Turn off the volume limiting.
Step 56 Press the Mode Select button and
select PSV from the Mode Select panel, but do
not press the confirm button. Advance to Step 57.