Page 227 of 292
Alarm 5. 101 System Fail (Memory Checksum Error)
Alarm message.................................... Monitor EEPROM fail
Alarm sub message ............................ Monitor checksum Fail
Alarm code........................................... 101
Priority of alarm................................... 5
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
At power up flow data in EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 6. 102 System Fail (Memory Checksum Error)
Alarm message.................................... Monitor EEPROM fail
Alarm sub message ............................ Monitor checksum Fail
Alarm code........................................... 102
Priority of alarm................................... 6
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
At power up oxygen data in EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 7. 103 System Fail (Memory Checksum Error)
Alarm message.................................... Monitor EEPROM fail
Alarm sub message ............................ Monitor checksum Fail
Alarm code........................................... 103
Priority of alarm................................... 7
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
At power up pressure offset data in EEPROM corrupt.
Action: Remove ventilator from service.