Alarm 8. 104 System Fail (Memory Checksum Error)
Alarm message ....................................Monitor EEPROM fail
Alarm sub message .............................Monitor checksum Fail
Alarm code ...........................................104
Priority of alarm ...................................8
Monitor mode .......................................All
Can alarm be muted ............................No
Sounder Priority...................................High
Alarm Description:
At power up pressure gain data in EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 9. 105 System Fail (Memory Checksum Error)
Alarm message ....................................Monitor EEPROM fail
Alarm sub message .............................Monitor checksum Fail
Alarm code ...........................................105
Priority of alarm ...................................9
Monitor mode .......................................All
Can alarm be muted ............................No
Sounder Priority...................................High
Alarm Description:
At power up pressure time constant data in EEPROM corrupt.
Action: Remove ventilator from service.
Alarm 10. Controller Failure
Alarm message ....................................Controller Failure
Alarm sub message .............................Controller has reset unexpectedly
Priority of alarm ...................................10
Monitor mode .......................................All
Can alarm be muted ............................No
Sounder Priority...................................High
Alarm Description:
If a command could not be sent to the alarm generator this message is produced.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.