Alarm 20. Unexpected Drop in Max P
Alarm message ....................................Unexpected drop in max P
Alarm sub message .............................Press autoset to adjust HFO alarms to
new pressures
Priority of alarm ...................................20
Monitor mode .......................................HFO Only
Can alarm be muted ............................Yes
Sounder Priority...................................High
Alarm Description:
When the ventilator set alarm threshold of 5mbar below the max pressure is crossed this
alarm is generated.
Action: Check patient. Press autoset or adjust ventilation parameters.
Check patient circuit & water trap.
Alarm 21. Unexpected Rise in Min P
Alarm message ....................................Unexpected rise in min P
Alarm sub message .............................Press autoset to adjust HFO alarms to
new pressures
Priority of alarm ...................................21
Monitor mode .......................................HFO Only
Can alarm be muted ............................Yes
Sounder Priority...................................High
Alarm Description:
When the ventilator set alarm threshold of 5mbar above the min pressure is crossed this
alarm is generated.
Action: Check patient. Press autoset or adjust ventilation parameters.
Alarm 22. Unexpected Drop in Min P
Alarm message ....................................Unexpected drop in min P
Alarm sub message .............................Press autoset to adjust HFO alarms to
new pressures
Priority of alarm ...................................22
Monitor mode .......................................HFO Only
Can alarm be muted ............................Yes
Sounder Priority...................................High
Alarm Description:
When the ventilator set alarm threshold of 5mbar below the min pressure is crossed this
alarm is generated.
Action: Check patient. Press autoset or adjust ventilation parameters.