Alarm 23. No Gas
Alarm message.................................... No gas
Alarm sub message ............................ Connect ventilator to gas
Priority of alarm................................... 23
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
If no gas is detected by the ventilator this alarm is generated.
Action: Connect gas supplies to ventilator. If generated whilst
connected to a patient, remove patient to alternative form of
ventilation.
Alarm 24. No O
2
Supply
Alarm message.................................... No O2 supply
Alarm sub message ............................ Connect oxygen supply to ventilator
Priority of alarm................................... 24
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
If no O
2
is detected by the ventilator this alarm is generated.
Action: Connect gas supplies to ventilator. If generated whilst
connected to a patient, remove patient to alternative form of
ventilation.
Alarm 25. No Air Supply
Alarm message.................................... No air supply
Alarm sub message ............................ Connect air supply to ventilator
Priority of alarm................................... 25
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
If no Air is detected by the ventilator this alarm is generated.
Action: Connect gas supplies to ventilator. If generated whilst
connected to a patient, remove patient to alternative form of
ventilation.