Page 234 of 292
Alarm 26. Battery Fault
Alarm message ....................................Battery fault
Alarm sub message .............................Internal battery not detected
Priority of alarm ...................................26
Monitor mode .......................................All
Can alarm be muted ............................No
Sounder Priority...................................High
Alarm Description:
The battery can not be detected by the ventilators monitoring system
Action: Remove ventilator from service. The ventilator will continue to
ventilate on mains power if connected to a patient.
Alarm 27. Battery Low
Alarm message ....................................Battery Low
Alarm sub message .............................Remove patient from ventilator
Priority of alarm ...................................27
Monitor mode .......................................All
Can alarm be muted ............................No
Sounder Priority...................................High
Alarm Description:
The battery low signal from the power supply sub-system is monitored while the system is
running on battery. When the signal is asserted an alarm indicating imminent power failure
is generated.
Action: Reconnect power mains power if available. If not available
remove patient to alternative form of ventilation immediately.
Alarm 28. Blocked Fresh Gas
Alarm message ....................................Blocked fresh Gas
Alarm sub message .............................Fresh gas supply to patient may be
blocked
Priority of alarm ...................................28
Monitor mode .......................................All
Can alarm be muted ............................No
Sounder Priority...................................High
Alarm Description:
If the fresh gas supply is blocked this alarm is generated.
Action: Check patient circuit.