EasyManua.ls Logo

SLE SLE5000 - Page 235

SLE SLE5000
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 235 of 292
Alarm 29. Leaking Fresh Gas
Alarm message.................................... Leaking fresh gas
Alarm sub message ............................ Fresh gas supply to patient may be
leaking
Priority of alarm................................... 29
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
If the fresh gas supply is detected to be leaking this alarm is generated.
Action: Check patient circuit.
Alarm 30. System Fail (Serial Communication Error)
Alarm message.................................... System fail
Alarm sub message ............................ Monitor error (comms)
Priority of alarm................................... 30
Monitor mode....................................... Flow
Can alarm be muted ........................... Yes
Sounder Priority .................................. High
Alarm Description:
Errors detected within the flow monitoring device.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.
Alarm 31. System Fail (Isolate System Error)
Alarm message.................................... System Fail
Alarm sub message ............................ Monitor isolated system fail
Priority of alarm................................... 31
Monitor mode....................................... Flow
Can alarm be muted ........................... Yes
Sounder Priority .................................. High
Alarm Description:
Errors detected within the flow monitoring device.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.

Table of Contents

Related product manuals