Page 237 of 292
Alarm 35. Connect Flow Sensor
Alarm message.................................... Connect flow sensor
Alarm sub message ............................ Flow sensor is not connected
Priority of alarm................................... 35
Monitor mode....................................... Flow
Can alarm be muted ........................... Yes
Sounder Priority .................................. Medium
Alarm Description:
If the sensor is not connected or both wires have been broken this alarm is generated.
Action: Connect flow sensor, if sensor already in situ replace and
discard the defective flow sensor.
Alarm 36. Cannot Calibrate Flow
Alarm message.................................... Flow calibration fail
Alarm sub message ............................ Unable to calibrate flow sensor
Priority of alarm................................... 36
Monitor mode....................................... Flow
Can alarm be muted ........................... Yes
Sounder Priority .................................. Medium
Alarm Description:
If calibrating flow fails for any reason this alarm is generated.
Action: Replace flow sensor.
Alarm 37. Calibrate Flow Sensor
Alarm message.................................... Calibrate flow sensor
Alarm sub message ............................ Calibrate flow sensor now
Priority of alarm................................... 37
Monitor mode....................................... Flow
Can alarm be muted ........................... Yes
Sounder Priority .................................. Medium
Alarm Description:
The above alarm is generated whenever the ventilator is turned ON or upon the re-
connection of the flow sensor.
Action: Carry out flow sensor calibration routine.