Alarm 38. User Interface Failure
Alarm message ....................................User interface failure
Alarm sub message .............................User interface has reset unexpectedly
Priority of alarm ...................................38
Monitor mode .......................................All
Can alarm be muted ............................Yes
Sounder Priority...................................High
Alarm Description:
When a hardware reset has taken place within the ventilator this alarm is generated.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.
Alarm 39. High Min. Volume
Alarm message ....................................High min. volume
Alarm sub message .............................Minute volume high threshold exceeded
Priority of alarm ...................................39
Monitor mode .......................................Flow
Can alarm be muted ............................Yes
Sounder Priority...................................High
Alarm Description:
Minute volumes that are higher than the user-selected thresholds generate this alarm.
Action: Check patient. Adjust ventilation parameters or alarm threshold.
Alarm 40. Patient Leak
Alarm message ....................................High patient leak
Alarm sub message .............................Check patient circuit
Priority of alarm ...................................40
Monitor mode .......................................Flow
Can alarm be muted ............................Yes
Sounder Priority...................................High
Alarm Description:
If the calculated patient leak is over user set threshold this alarm is generated.
Action: Check patient. Check patient circuit and ET tube. Adjust alarm
threshold.