Alarm 47. Main Power Failure
Alarm message.................................... Main power fail
Alarm sub message ............................ Running on internal battery
Priority of alarm................................... 47
Monitor mode....................................... All
Can alarm be muted ........................... Yes
Sounder Priority .................................. High
Alarm Description:
If the mains supply fails this alarm is generated.
Action: Check mains connection. Prepare alternative form of ventilation
whilst running on battery.
Alarm 48. Fresh Gas Solenoid Fail
Alarm message.................................... Fresh gas solenoid fail
Alarm sub message ............................ Remove ventilator from service
Priority of alarm................................... 48
Monitor mode....................................... All
Can alarm be muted ........................... Yes
Sounder Priority .................................. High
Alarm Description:
When the ventilator pressure transducer detects no gas from the solenoid this alarm is
generated.
Action: Remove patient to alternative form of ventilation immediately,
then remove ventilator from service.
Alarm 49. Oxygen Cell Discon. (Disconnected)
Alarm message.................................... Oxygen cell discon.
Alarm sub message ............................ The oxygen cell is not connected
Priority of alarm................................... 49
Monitor mode....................................... All
Can alarm be muted ........................... Yes
Sounder Priority .................................. Medium
Alarm Description:
A disconnection of the oxygen cell will generate this alarm.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.