Alarm 50. Calibrate Oxygen Cell
Alarm message ....................................Calibrate oxygen cell
Alarm sub message .............................The oxygen cell needs calibrating
Priority of alarm ...................................50
Monitor mode .......................................All
Can alarm be muted ............................Yes
Sounder Priority...................................Medium
Alarm Description:
If at any time the measured oxygen is >100% this alarm will be generated.
Action: Recalibrate O
2
sensor. If alarm re-appears remove patient to
alternative form of ventilation, then remove ventilator from
service.
Alarm 51. Oxygen Cell Exhausted
Alarm message ....................................Oxygen cell exhausted
Alarm sub message .............................A new oxygen cell is required
Priority of alarm ...................................51
Monitor mode .......................................All
Can alarm be muted ............................Yes
Sounder Priority...................................Medium
Alarm Description:
If the oxygen cell could not be calibrated during calibration this alarm will be generated.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.
Alarm 52. Oxygen Calibration Failure
Alarm message ....................................O
2
calibration fail
Alarm sub message .............................Check O
2
supply
Priority of alarm ...................................52
Monitor mode .......................................All
Can alarm be muted ............................Yes
Sounder Priority...................................Medium
Alarm Description:
If during 100% oxygen calibration the measured oxygen level is below a certain level this
alarm is generated.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.