Alarm 53. High Oxygen Level
Alarm message.................................... High oxygen level
Alarm sub message ............................ The O
2
is higher than desired
Priority of alarm................................... 53
Monitor mode....................................... All
Can alarm be muted ........................... Yes
Sounder Priority .................................. High
Alarm Description:
If the delivered oxygen differs from the user selected level by more than 5% the above
alarm is generated.
Action: Recalibrate O
2
sensor. If alarm re-appears remove patient to
alternative form of ventilation, then remove ventilator from
service.
Alarm 54. Low Oxygen Level
Alarm message.................................... Low oxygen level
Alarm sub message ............................ The O
2
is lower than desired
Priority of alarm................................... 54
Monitor mode....................................... All
Can alarm be muted ........................... Yes
Sounder Priority .................................. High
Alarm Description:
If the delivered oxygen differs from the user selected level by more than 5% the above
alarm is generated.
Action: Recalibrate O
2
sensor. If alarm re-appears remove patient to
alternative form of ventilation, then remove ventilator from
service.
Alarm 55. Monitor/Display Comms Fail
Alarm message.................................... Monitor/display comms fail
Alarm sub message ............................ Call engineer
Priority of alarm................................... 55
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
When the ventilator detects a communications failure between the contoller and the
monitor/display system this alarm is generated.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.