HRX-OM-X037
Chapter 5 Display and Setting of Various Functions
HRL Series 5.4 Description of the Screen
Table 5.4-11 TEMP READY signal setting
Upper/lower
temperature
limit
Set the temperature range of
the circulating fluid SP.
+0.1 ºC to +10.0 ºC
+1.0 ºC
Set the temperature range of
the circulating fluid SP.
-0.1 ºC to -10.0 ºC
-1.0 ºC
Stable
temperature
time/extra
time
“TEMP READY” signal is output
when the set time has passed
after the circulating fluid
temperature rises/drops within
the range of (1) “Upper/Lower
temperature limit.”
10sec to 9999sec
180sec
“TEMP READYˮ is cleared
when the set time has passed
after the circulating fluid
temperature rises/drops outside
the range of (1) “Upper/Lower
temperature limit” in the “TEMP
READY” condition.
Operation continues during the
alarm
Operation stops during alarm
Start time for
AL16 alarm
monitoring
Monitoring of “AL16: TEMP
READY alarm” starts when the
set time has passed after the
start of operation.
By default
TEMP READY signal (AL16) setting