TITANUS
MICRO·SENS
®
Design
06/13
67
In conformity with EN 54-20
or ISO 7240-20
Triggering threshold 10 % 20 % 40 % 50 %
Sensitivity Very high High Medium Low
TIP
It is recommended to always select the greatest possible level which is
permissible according to national standards.
The device’s air flow monitoring enables the system to detect both pipe
breakages outside the device and sudden obstruction in individual aspiration
apertures (e.g. in the event of sabotage to the pipe system). The dynamic air
flow sensor system has been activated via the diagnostics software; the fol-
lowing modifications have to be regarded.
The air flow monitoring may only be set, if:
▪ Project planning according to “Individual aperture monitoring” has been
carried out “(see Chap. 5.3.1 “Pipe project planning individual aperture
monitoring“),
▪ the air flow sensor has been compensated depending on the air pres-
sure (see Chap. 8.1.2 “Air pressure dependent air flow compensation“)
und
▪ No large air flow fluctuations occur.
The same air pressure must be present throughout the sampling pipe.
NOTICE
If the air sampling smoke detection system and pipe system are in areas
with different air pressure, the air sampled by the TITANUS
®
should be re-
circulated in the pipe system pressure area (see Chapter “Air recircula-
tion”).
Dynamic air flow senso
system
Limitations
ir pressure differences