EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1100 Series

Agilent Technologies 1100 Series Service Manual

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
94 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
System Pressure Abnormally High
The current system pressure is significantly above the typical system pressure
produced by this method with this column.
Table 7 System Pressure Abnormally High: Possible causes and actions
Possible Causes Suggested Actions Notes
The analytical column has become
plugged
Replace the column. Otherwise, backflush the
column or replace the column inlet frit.
The filter in front (upstream) of the
EMPV has become plugged
Perform a pump purge at 1000 µl/min using pure
water. During the purge, check system pressure. If
pressure is >10 bar, replace the EMPV filter.
I
A component in the micro-sampler
has become plugged. This could be
the sample loop, needle, needle
seat assembly or injection valve
ports.
Using the sampler maintenance positions, switch
the sampler injection valve from mainpass to
bypass. If pressure is significantly reduced:
Backflush or change the needle seat assembly.
Replace the needle.
Backflush or replace the sample loop capillary.
Replace the injection valve rotor seal.
Clean the stator head with acetone, and make
sure the stator head ports are free of particles.
For severe capillary tube
plugs, acetone is a good
backflushing solvent.
A capillary before or after the
sampler in the system has become
plugged, broken, crushed by a
module cover or overtightened.
Refer to the system flow diagram. One at a time,
disconnect the capillaries in the following order.
When the defective capillary is found, it may be
backflushed with acetone, or replaced.
EMPV-to-flow sensor capillary
flow sensor-to-sampler injection valve capillary
sampler injection valve-to-column inlet
capillary
flow cell assembly (includes inlet and outlet
capillaries)

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals