1100 Series Nano Pump Service Manual 31
Troubleshooting and Test Functions 2
Leak
A leak was detected in the nano pump.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the CSM board.
Probable Causes
Loose fittings.
Broken capillary.
Loose or leaking active inlet valve, outlet ball valve, or EMPV.
Defective pump seals.
Suggested Actions
Ensure all fittings are tight.
Exchange defective capillaries.
Ensure pump components are seated correctly. If there are still signs of a
leak, exchange the appropriate seal (active inlet valve, outlet ball valve).
Exchange the pump seals.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals