1100 Series Nano Pump Service Manual 33
Troubleshooting and Test Functions 2
Leak Sensor Short
The leak sensor in the nano pump has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Probable Causes
Defective leak sensor.
Leak sensor incorrectly routed, being pinched by a metal component.
Suggested Actions
Exchange the leak sensor.