1100 Series Nano Pump Service Manual 39
Troubleshooting and Test Functions 2
Zero Solvent Counter
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings
at the ChemStation (revision 5.xx and higher). If the volume level in the bottle
falls below the specified value the error message appears when the feature is
configured accordingly.
Probable Causes
Volume in bottle below specified volume.
Incorrect setting of limit.
Suggested Actions
Refill bottles and reset solvent counters.