110 1100 Series Nano Pump Service Manual
3 Repairing the Pump
2 Before inserting the new valve cartridge clean the area in the actuator
assembly. Fill a syringe with alcohol and flush the cartridge area
thoroughly.
3 Insert a new valve cartridge into the actuator assembly. Make sure the valve
cartridge is fully inserted into the actuator assembly.
Replacing the Active Inlet Valve
1 Insert the new valve into the pump head. Using the 14 mm wrench turn the
nut until hand tight.
2 Position the valve that the solvent inlet tube connection points towards the
front.
3 Using the 14 mm wrench tighten the nut by turning the valve in its final
position (not more than a quarter turn). Do not overtighten the valve. The
solvent inlet tube connection should point to the right corner of the pump
head.
4 Reconnect the inlet tube and the active inlet valve cable to the connector at
the Z-panel.
5 After an exchange of the valve cartridge it may take several mL of pumping
with the solvent used in the current application, before the flow stabilizes
at A%-ripple as low as it used to be, when the system was still working
properly.
NOTE
Make sure you are in normal mode.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals