32 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Leak Sensor Open
The leak sensor in the nano pump has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes
Leak sensor not connected to the CSM board.
Defective leak sensor.
Leak sensor incorrectly routed, being pinched by a metal component.
Suggested Actions
Ensure the leak sensor is connected correctly.
Exchange the leak sensor.