1100 Series Nano Pump Service Manual 47
Troubleshooting and Test Functions 2
Inlet-Valve Fuse
Inlet-Valve Fuse 0: Pump channel A
Inlet-Valve Fuse 1: Pump channel B
One of the active-inlet valves in the nano pump has drawn excessive current
causing the inlet-valve electronic fuse to open.
Probable Causes
Defective active inlet valve.
Defective connection cable (front panel to CSM board).
Defective CSM board.
Suggested Actions
Restart the nano pump. If the error message appears again, exchange the
active inlet valve.
Exchange the connection cable.
Exchange the CSM board.