226 1100 Series Nano Pump Service Manual
5 Introduction to the Nano Pump
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
Forced Cold Start Settings
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
To return to normal operation, set switches back to your GPIB or RS 232
configuration settings.
Stay-Resident Settings
Firmware update procedures may require this mode in case of firmware
loading errors.
1 0 Odd Parity
1 1 Even Parity
Table 59 Parity Settings, continued
CAUTION
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are diagnose and repair log books which will not be erased.
Table 60 Forced Cold Start Settings
Mode Select12345678
TEST/BOOT11000001