EasyManuals Logo

Agilent Technologies 1100 Series Service Manual

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #238 background imageLoading...
Page #238 background image
226 1100 Series Nano Pump Service Manual
5 Introduction to the Nano Pump
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
Forced Cold Start Settings
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
To return to normal operation, set switches back to your GPIB or RS 232
configuration settings.
Stay-Resident Settings
Firmware update procedures may require this mode in case of firmware
loading errors.
1 0 Odd Parity
1 1 Even Parity
Table 59 Parity Settings, continued
CAUTION
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are diagnose and repair log books which will not be erased.
Table 60 Forced Cold Start Settings
Mode Select12345678
TEST/BOOT11000001

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals