EasyManua.ls Logo

Agilent Technologies 1100 Series - Pressure below Lower Limit

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1100 Series Nano Pump Service Manual 41
Troubleshooting and Test Functions 2
Pressure Below Lower Limit
The system pressure has fallen below the lower pressure limit.
Probable Causes
Lower pressure limit set too high.
Air bubbles in the mobile phase.
Leak.
Defective damper.
Defective CSM board.
Suggested Actions
Ensure the lower pressure limit is set to a value suitable for the analysis.
Ensure solvents are degassed. Purge the nano pump.
Ensure solvent inlet filters are not blocked.
Inspect the pump head, capillaries and fittings for signs of a leak.
Purge the nano pump. Run a pressure test to determine whether the seals
or other pump components are defective.
Exchange the damper.
Exchange the CSM board.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals