52 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Inlet-Valve Missing
Inlet-Valve Missing: Pump channel A
B: Inlet-Valve Missing: Pump channel B
The active-inlet valve in the nano pump is missing or defective.
The processor checks the presence of the active-inlet valve connector every
2 seconds. If the connector is not detected by the processor, the error message
is generated.
Probable Causes
Disconnected or defective cable.
Disconnected or defective connection cable (front panel to CSPM board).
Defective active inlet valve.
Suggested Actions
Ensure the pins of the active inlet valve connector are not damaged. Ensure
the connector is seated securely.
Ensure the connection cable is seated correctly. Exchange the cable if
defective.
Exchange the active inlet valve.