248 1100 Series Nano Pump Service Manual
6 Control Module Screens for the Nano Pump
Screens available from the Records screen
Records screen Use the Esc-key until you receive Views on the F5-key. Choose System from
the pull-down menu. Use the F4-key (Records) and select the pump. Errors
are reported either into the F2 (System Log) or F3 (Error Log).
EMF
Early Mainte-
nance Feedback)
Use the F1-key (EMF) to set EMF Limits. Choose menu item Setup Limits to
set the amount of pumped solvent or seal wear, when you want to receive a
warning for exceeded limits.
Press F7 (Reset) to reset the counters after exchanging parts that are subject
to wear. Changes must be activated with F6 (Done).