56 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Leak Sensor Missing
Probable Causes
Disconnected or defective cable.
Defective leak sensor.
Suggested Actions
Ensure the connection cable is seated correctly.
Exchange the leak sensor.