1100 Series Nano Pump Service Manual 63
Troubleshooting and Test Functions 2
Initialization Failed
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The nano pump failed to initialize successfully within the maximum time
window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Probable Causes
Blocked active inlet valve.
Defective pump drive assembly.
Defective CSM board.
Suggested Actions
Exchange the active inlet valve.
Exchange the pump drive assembly.
Exchange the CSM board.