1100 Series Nano Pump Service Manual 95
Troubleshooting and Test Functions 2
EMPV failed to initialize (micro mode only)
An attempt to pump in the micro mode has resulted in either an EMPV 
Initialization Failed error message, or a permanent EMPV Initialization not 
ready message.
Make sure the system pressure is higher than 20 bar.
Table 8 EMPV failed to initialize: Possible Causes and Suggested Actions
Possible Causes  Suggested Actions  Notes
The no-flow pressure of the system 
is higher than 10 bar.
• Set the flow to zero, and disconnect the blue 
flexible capillary going from the damper to the 
mixer The system pressure reading should be 
close to zero bar. 
• If the system pressure reading is higher than 4 
bar, call Agilent service, or refer to the Nano Pump 
Service Manual.
This problem typically causes 
a permanent EMPV 
Initialization not ready 
message.
The inlet to the EMPV has been 
blocked, or partially restricted. The 
EMPV cannot take in sufficient 
flow to deliver the correct flow 
output. The EMPV initialization 
routine cannot be done within the 
required 2-minute period. 
• Make sure the solvent channels are well purged.
• Check the EMPV filter. Perform a pump purge at 
1000 µl/min using pure water. During the purge, 
check system pressure. If pressure is >10 bar, 
replace the filter in front of the EMPV. 
• Check the flow path from the damper outlet to the 
EMPV inlet for plugs or restrictions.
• Check the EMPV-to-flow sensor capillary for a 
total plug or partial restriction. Replace the 
capillary, or backflush the capillary with acetone.
• Replace the EMPV assembly (G1361-60000). Call 
Agilent service or refer to the Nano Pump Service 
Manual "Exchanging the EMPV Assembly" on 
page 117.
This problem typically causes 
an EMPV Initialization Failed 
error message