EasyManuals Logo

Agilent Technologies 1100 Series Service Manual

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
1100 Series Nano Pump Service Manual 61
Troubleshooting and Test Functions 2
Index Missing
Index Missing: Pump channel A
B: Index Missing: Pump channel B
The encoder index position in the nano pump was not found during
initialization.
During initialization, the first plunger is moved to the mechanical stop. After
reaching the mechanical stop, the plunger reverses direction until the encoder
index position is reached. If the index position is not recognized within a
defined time, the error message is generated.
Probable Causes
Disconnected or defective encoder cable.
Defective pump drive assembly.
Suggested Actions
Ensure the encoder cable are not damaged or dirty. Make sure the cables
are connected securely to the CSM board.
Exchange the pump drive assembly.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals