46 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Valve Fuse
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent-selection valves in the nano pump has drawn excessive
current causing the selection-valve electronic fuse to open.
Probable Causes
Defective solvent selection valve.
Defective connection cable (front panel to CSM board).
Defective CSM board.
Suggested Actions
Restart the nano pump. If the error message appears again, exchange the
solvent selection valve.
Exchange the connection cable.
Exchange the CSM board.