1100 Series Nano Pump Service Manual 13
Installing the Pump 1
Installing the Nano Pump
1 Place the pump on top of the micro-sampler. Make sure that the two
modules are interlocked correctly.
2 Connect the power cable to the power socket at the rear of the pump. Do not
connect the power cable to power until you have finished the hardware
installation of all modules in the stack.
3 Connect the free end of the Can-bus cable from the micro-sampler to one of
the Can-bus connectors at the rear of the pump.
Figure 5 Rear of Nano Pump
GPIB
Security Lever
Configuration switch
Power Plug
RS232
Remote
Analog output
Slot for interface board
CAN-Bus