EasyManua.ls Logo

Agilent Technologies 1100 Series - Missing Pressure Reading

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Missing Pressure Reading
The pressure readings read by the pump ADC (analog-digital converter) are
missing.
The ADC reads the pressure readings from the damper every 1ms. If the
readings are missing for longer than 10 seconds, the error message is
generated.
Probable Causes
Damper not connected.
Defective damper.
Defective CSM board.
Suggested Actions
Ensure the damper connector is clean, and seated correctly.
Exchange the damper.
Exchange the CSM board.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals