28 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The nano pump continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
Leak detected in another module with a CAN connection to the system.
Leak detected in an external instrument with a remote connection to the
system.
Shut-down in an external instrument with a remote connection to the
system.
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
Fix the leak in the external instrument before restarting the nano pump.
Check external instruments for a shut-down condition.
Check the vacuum degasser for an error condition. Refer to the Reference
Manual for the Agilent 1100 Series vacuum degasser.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals