114 1100 Series Nano Pump Service Manual
3 Repairing the Pump
Installing the Manual Purge Valve
Installing the Purge Valve
1 Switch off the pump module.
2 Disconnect the capillary from the mixer to the filter at the filter inlet.
3 Connect the capillary end into the purge valve holder.
4 Adapt the purge valve holder on the pump head of channel A.
5 Use the screw and the washer to fix the purge valve holder.
6 Insert the purge valve assembly into the purge valve holder and locate the
outlet and waste outlet as shown below.
7 Use the 14 mm wrench to tighten the purge valve assembly.
8 Connect the waste tube from the EMPV assembly to the waste outlet of the
purge valve.
9 Connect the pump outlet capillary into the outlet of the purge valve.
Frequency When performing a leak test for verifying pump tightness
When using the nano pump permanently in normal mode
Tools required Wrench 1/4 inch
Wrench 14 mm
Wrench 4 mm
Parts required Purge valve assy G1311-60009
Adapter G1312-23200
Screw 0515-0175
Washer 2190-0586

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals