40 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Pressure Above Upper Limit
The system pressure has exceeded the upper pressure limit.
Probable Causes
Upper pressure limit set too low.
Blockage in the flowpath (after the damper).
Defective damper.
Defective CSM board.
Suggested Actions
Ensure the upper pressure limit is set to a value suitable for the analysis.
Check for blockage in the flowpath.
Exchange the damper.
Exchange the CSM board.